Testing of laser ablation ion source for JYFLTRAP

In this work, we have constructed and tested a laser ablation ion source for JYFLTRAP Penning trap at the IGISOL (Ion Guide Isotope Separator On-line) facility. The calibration of the Penning trap parameters requires reference ions or ion clusters that have well-known masses with relatively small ma...

Täydet tiedot

Bibliografiset tiedot
Päätekijä: Poleshchuk, Kateryna
Muut tekijät: Matemaattis-luonnontieteellinen tiedekunta, Faculty of Sciences, Fysiikan laitos, Department of Physics, University of Jyväskylä, Jyväskylän yliopisto
Aineistotyyppi: Pro gradu
Kieli:eng
Julkaistu: 2015
Aiheet:
Linkit: https://jyx.jyu.fi/handle/123456789/48660
Kuvaus
Yhteenveto:In this work, we have constructed and tested a laser ablation ion source for JYFLTRAP Penning trap at the IGISOL (Ion Guide Isotope Separator On-line) facility. The calibration of the Penning trap parameters requires reference ions or ion clusters that have well-known masses with relatively small mass uncertainty. These ions and ion clusters can be created with certain solid targets, which contain large amounts of isotopes of reference masses and by using a laser for target ablation. In this work simulations of ion source parameters, construction, testing and improving a laser ablation ion source setup have been done. Targets with relatively large abundances of 63Cu and 65Cu, 85Rb, 87Rb, 115In, 120Sn have been used for the laser ablation. Some of these ions were successfully created and measured with the Penning trap.